Related Experiment Videos

Antimicrobial susceptibility testing of Haemophilus influenzae. Improvement of accuracy of the disc diffusion test

S Ringertz1, B Olsson-Liljequist, G Kronvall

  • 1Department of Clinical Microbiology, Karolinska Institute, Karolinska Hospital, Stockholm, Sweden.

Insights

A 1986 study revealed inaccurate antimicrobial susceptibility testing for Haemophilus influenzae in Sweden. A new standardized method and breakpoint system implemented in 1988 significantly improved accuracy, virtually eliminating interpretive errors.

Area of Science:

  • Medical Microbiology
  • Clinical Diagnostics
  • Pharmaceutical Science

Background:

  • The disc diffusion method is crucial for antimicrobial susceptibility testing.
  • In 1986, a national quality control study in Sweden identified unacceptably low accuracy in interpreting Haemophilus influenzae susceptibility results.
  • This highlighted a need for improved standardization and interpretive criteria.

Purpose of the Study:

  • To develop and evaluate a new standardized method for antimicrobial susceptibility testing of Haemophilus influenzae.
  • To establish a revised system for setting interpretive zone breakpoints.
  • To improve the accuracy of susceptibility interpretations in clinical laboratories.

Main Methods:

  • A national quality control study was conducted in 1986.
  • A new standardized disc diffusion method was developed.
  • Interpretive zone breakpoints were recalculated using MIC50 values and zone histograms from five reference laboratories.
  • A follow-up quality control study was performed in 1988 to assess the impact of the new methods.

Main Results:

  • The initial 1986 study showed unacceptably low accuracy in susceptibility interpretations.
  • The newly developed standardized method and breakpoint system significantly improved accuracy in the 1988 follow-up study.
  • Interpretive errors were virtually eliminated for most antibiotics, with specific adjustments for doxycycline improving accuracy in all laboratories.

Conclusions:

  • The revised standardized method and breakpoint setting strategy effectively improved the accuracy of antimicrobial susceptibility testing for Haemophilus influenzae.
  • The new system allows for correct categorization of bacterial susceptibility and detection of reduced susceptibility.
  • Further refinement, including laboratory-specific breakpoints for certain antibiotics like doxycycline, ensures robust diagnostic performance.

Related Concept Videos