Related Experiment Video
Updated: May 22, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Auscultatory versus oscillometric measurement of blood pressure in octogenarians
Jens-Ulrik Rosholm1, Sidsel Arnspang, Lars Matzen
1Department of Geriatrics, Odense University Hospital, Denmark.
Background:
Auscultatory measurement using a sphygmomanometer has been the predominant method for clinical estimation of blood pressure, but it is now rapidly being replaced by oscillometric measurement.
Objective:
To compare blood pressure by auscultatory and oscillometric measurements in patients ≥ 80 years.
Method:
100 patients had blood pressure measured by auscultation with a sphygmomanometer and by an electronic device using the oscillometric method. For each patient the mean of two blood pressures with each method measured within 15 min were compared.
Results:
The mean age of participants was 85.8 years; 55.8% were women. The correlation coefficient for systolic blood pressure was 0.88 and for diastolic 0.79. Differences between auscultatory and oscillometric values were less than 10 mmHg in 70.6% of systolic blood pressures and in 83.2% for diastolic. Arrhythmia and hypertension did not influence the results, and there was no correlation between the magnitude of the differences and the level of blood pressure.
Conclusion:
Agreement between oscillometric and auscultatory measurements of blood pressure in octogenarians was found to be less than required by validation protocols. However, semi-automatic equipment, which is observer-independent, may be used even in the very elderly, particularly if multiple readings are performed.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...