Related Experiment Video
Updated: May 22, 2026

Ratiometric Imaging of Extracellular pH in Dental Biofilms
Published on: March 9, 2016
Error analysis of ratiometric imaging of extracellular pH in a window chamber model
Yuxiang Lin1, Tzu-Yu Wu, Arthur F Gmitro
1University of Arizona, College of Optical Sciences, 1630 East University Boulevard, Tuscon, Arizona 85721, USA.
Abstract:
Ratiometric fluorescence-imaging technique is commonly used to measure extracellular pH in tumors and surrounding tissue within a dorsal skin-fold window chamber. Using a pH-sensitive fluorophore such as carboxy SNARF-1 one can measure pH distributions with high precision. However, it is often observed that the measured pH is lower than expected, with a bias that varies from one image to another. A comprehensive analysis of possible error sources is presented. These error sources include photon noise, estimator bias, instrument errors, temperature, and calibration errors from biological factors.