Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Comparison of 2 portable respiratory gas analysers
P M Leprêtre1, T Weissland, C Paton
1Laboratoire de Recherche Adaptations Physiologiques à l'Exercice et -Réadaptation à l'Effort Université de Picardie Jules Verne, AMIENS, France. pierre-marie.lepretre@u-picardie.fr
Abstract:
The aim of this study was to compare respiratory measures taken simultaneously using the Cosmed K4b(2) and Cortex Metamax II portable metabolic systems. 10 trained male cyclists performed a graded exercise cycle test to exhaustion (40w.3 min (- 1)) under standardized conditions. The measured respiratory variables were significantly correlated between both devices: r=0.97 and 0.98 (n=10, p<0.01) for oxygen uptake ( V˙O(2)) and ventilation (VE), respectively. Further Bland and Altman plots revealed a good level of agreement for measures of V˙O(2) expressed in mL.min (- 1) [- 670 to 486] (mean bias of - 91.7, i. e., - 3.1%) or in mL.min (- 1).kg (- 1) [- 7.3 to 9.0] (mean bias of 0.85, i. e., 2.3%) and V(E) [- 23.1 to 18.2] (L.min (- 1), mean bias of - 2.4, i. e. - 4.1%). However, poor agreement was found for measures of carbon dioxide (VCO(2), mL.min (- 1)) [- 280 to 1 394] (mean bias of 671 i.e., 20.3%). VCO(2) at maximal exercise intensity was also significantly (p<0.01) greater in the Cortex compared to the Cosmed system. The higher measured CO(2) concentrations led to significantly (p<0.01) higher calculated respiratory exchange ratio (RER) values with the Cortex device. In conclusion, there was satisfactory agreement between the Cosmed K4b(2) and Cortex Metamax II systems for most respiratory measures; however there was a poor level of agreement between VCO(2) and calculated RER measurements between the 2 systems.
More Related Videos
08:44Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
09:34Implementation of Portable Emissions Measurement Systems (PEMS) for the Real-driving Emissions (RDE) Regulation in Europe
Published on: December 4, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...