Evaluations of absorbed dose ratio factor of Al2O3 dosemeter in radiotherapy photon beams using cavity theory
Jinhan Zhu1, Shaowen Chen, Lixin Chen
1School of Physics Science and Engineering, Sun Yat-Sen University, Guangzhou 510275, China.
Abstract:
The aim of the work was to evaluate the absorbed dose ratio factor f(md) of an Al(2)O(3) dosemeter to water in photon radiotherapy beams using cavity theory. Burlin theory was used for calculating of this ratio. The effective mass attenuation coefficient β was obtained by comparing Monte Carlo simulations in monoenergetic photon beams. The evaluations of the absorbed dose ratio factor f(md) were studied for Al(2)O(3) dosemeters of different sizes, which were placed at various depths of the water phantom in different radiation field sizes of Mohan's 6, 10 and 15-MV X-rays. Beyond the build-up region, the variation of f(md) increases by 0.25 % as the depth increases from 4 to 10 cm. The maximum variation due to different dosemeter sizes is 8.3 %. The difference in the f(md) due to different radiation field sizes is 1.5 %. The effect of the dosemeter size cannot be neglected. The difference in the f(md) due to the radiation field sizes of different beams would increase as the dosemeter size increases.