Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
Pulmonary venous anatomy imaging with low-dose, prospectively ECG-triggered, high-pitch 128-slice dual-source
Wai-ee Thai1, Bryan Wai, Kaity Lin
1Cardiac MR PET CT Program, Division of Cardiology and Department of Radiology and the Cardiac Arrhythmia Service, Massachusetts General Hospital, Harvard Medical School, Boston, MA 02114, USA.
Background:
The efforts to reduce radiation from cardiac computed tomography (CT) are essential. Using a prospectively triggered, high-pitch dual-source CT protocol, we aim to determine the radiation dose and image quality in patients undergoing pulmonary vein (PV) imaging.
Methods And Results:
In 94 patients (61±9 years; 71% male) who underwent 128-slice dual-source CT (pitch 3.4), radiation dose and image quality were assessed and compared between 69 patients with sinus rhythm and 25 patients with atrial fibrillation. Radiation dose was compared in a subset of 19 patients with prior retrospective or prospectively triggered CT PV scans without high pitch. In a subset of 18 patients with prior magnetic resonance imaging for PV assessment, PV anatomy and scan duration were compared with high-pitch CT. Using the high-pitch protocol, total effective radiation dose was 1.4 (1.3, 1.9) mSv, with no difference between sinus rhythm and atrial fibrillation (1.4 versus 1.5 mSv; P=0.22). No high-pitch CT scans were nondiagnostic or had poor image quality. Radiation dose was reduced with high-pitch (1.6 mSv) compared with standard protocols (19.3 mSv; P<0.0001). This radiation dose reduction was seen with sinus rhythm (1.5 versus 16.7 mSv; P<0.0001) but was more profound with atrial fibrillation (1.9 versus 27.7 mSv; P=0.039). There was excellent agreement of PV anatomy (κ 0.84; P<0.0001) and a shorter CT scan duration (6 minutes) compared with magnetic resonance imaging (41 minutes; P<0.0001).
Conclusions:
Using a high-pitch dual-source CT protocol, PV imaging can be performed with minimal radiation dose, short scan acquisition, and excellent image quality in patients with sinus rhythm or atrial fibrillation. This protocol highlights the success of new cardiac CT technology to minimize radiation exposure, giving clinicians a new low-dose imaging alternative to assess PV anatomy.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...