Related Experiment Video
Updated: May 22, 2026

Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
Development of normal-suction boundary control method based on inflow cannula pressure waveform for the undulation
Kohei Ishii1, Itsuro Saito, Takashi Isoyama
1Department of Biomedical Engineering, University of Tokyo, Bunkyo-ku, Tokyo, Japan. ishii@bme.gr.jp
Abstract:
It is desirable to obtain the maximum assist without suction in ventricular assist devices (VADs). However, high driving power of a VAD may cause severe ventricle suction that can induce arrhythmia, hemolysis, and pump damage. In this report, an appropriate VAD driving level that maximizes the assist effect without severe systolic suction was explored. The target driving level was set at the boundary between low driving power without suction and high driving power with frequent suction. In the boundary range, intermittent mild suction may occur. Driving power was regulated by the suction occurrence. The normal-suction boundary control method was evaluated in a female goat implanted with an undulation pump ventricular assist device (UPVAD). The UPVAD was driven in a semipulsatile mode with heartbeat synchronization control. Systolic driving power was adjusted using a normal-suction boundary control method developed for this study. We confirmed that driving power could be maintained in the boundary range. Occurrences of suction were evaluated using the suction ratio. We defined this ratio as the number of suction occurrences divided by the number of heartbeats. The suction ratio decreased by 70% when the normal-suction boundary control method was used.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)