High-definition computed tomography for coronary artery stents imaging compared with standard-definition 64-row
Wen Jie Yang1, Huan Zhang, Hua Xiao
1Department of radiology, Rui Jin Hospital, Medical School, Shanghai Jiao Tong, University, Shanghai, People's Republic of China.
Objective:
The evaluation of coronary stents by computed tomography (CT) remains difficult. We assessed the imaging performance of a high-definition CT scanner (HDCT) by comparing with a conventional 64-row standard-definition CT (SDCT).
Methods:
One hundred thirty-eight consecutive stented patients underwent coronary CT angiography, among whom 66 patients were examined by HDCT, and 72 patients by SDCT (LightSpeed VCT XT; GE Healthcare, Waukesha, Wis). The image quality score, the inner stent diameter (ISD), and the radiation dose were analyzed. All data were statistically tested by SPSS 13.0 software (SPSS Inc, Chicago, Ill).
Results:
In 72 patients examined using SDCT, 135 stents were detected; in 66 patients examined using HDCT, 119 stents were detected. The image quality score on HDCT was significantly better than that on SDCT (1.4 [SD, 0.7] vs 1.9 [SD, 0.8]). The ISD on HDCT was significantly higher than that on SDCT (1.8 [SD, 0.5] vs 1.6 [SD, 0.4]). There was no significant difference of either image quality score or ISD between the HDCT and SDCT groups in stents with 2.5-mm diameter. Images on HDCT showed significantly better image quality score and larger ISD than images on SDCT in 2.75-, 3-, and 3.5-mm stents. For patients examined by retrospective electrocardiogram-gated technique, the radiation dose on HDCT was significantly lower than that on SDCT (11.3 [SD, 2.9] vs 15.1 [SD, 3.8] mSv).
Conclusions:
High-definition CT scanner offered improved image quality and measurement accuracy for imaging coronary stents compared with conventional SDCT, providing higher spatial resolution and lower dose for evaluating coronary stents with 2.75- to 3.5-mm diameter.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...