Related Experiment Video
Updated: May 22, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Hip resurfacing with the Biomet Hybrid ReCap-Magnum system: 7-year results
1Midlands Orthopaedics, P.A., Columbia, SC, USA.
Abstract:
The purpose of this study was to report our clinical outcome of a large series of metal-on-metal hip resurfacing arthroplasty (HRA) using the hybrid Biomet ReCap-Magnum system. This is a single-designer surgeon series with an average of 5 ± 1 years. Seven hundred forty consecutive hybrid HRAs were performed in 653 patients. Kaplan-Meier survivorship with any revision as an end point was 96.4% at 7 years. Twenty-five (3.4%) cases were revised: 8 due to acetabular component loosening, 6 due to femoral neck facture, 4 due to failure of femoral component fixation, 2 due to deep infection, 2 due to adverse wear, 1 due to psoas tendonitis, 1 due to recurrent dislocation, and 1 due to unexplained pain. Biomet ReCap and Magnum HRA components with hybrid fixation methods showed excellent survivorship for a minimally selected young patient cohort at 7 years.