Investigations concerning the application of the cross-correlation method in cardiac output measurements
Maciej Gawlikowski1, Tadeusz Pustelny
1Foundation of Cardiac Surgery Development, Artificial Heart Lab,, Zabrze, Poland. maciej.gawlikowski@poczta.onet.pl
Background:
In spite of numerous non-invasive examinations the "gold clinical standard" of cardiac output measurements is the invasive pulmonary artery catheterization by means of the Swan-Ganz catheter and the application of the thermodilution method to estimate the blood flow. The results obtained by means of thermodilution are sensitive to many physical and biological disturbances. The unreliability of this method amounts to 20-45% and depends on the given variant of the method. Therefore some other method, more accurate and resistant to disturbances, was looked for. This paper presents a new approach to cardiac output measurements, based on cross-correlation signal analysis. The goal of investigations was to verify experimentally the application of the cross-correlation method of cardiac output measurements.
Results:
In 99.2% of the examined cases the extreme of the cross-correlation function was easy to be estimated by numerical algorithms. In 0,8% of the remaining cases (with a plateau region adjacent to the maximum point) numerical detection of the extreme was inaccurate. The typical unreliability of the investigated method amounted o 5.1% (9.8% in the worst case). Investigations performed on a physical model revealed that the unreliability of cardiac output measurements by means of the cross-correlation method is 3-5 times better than in the case of thermodilution.
Conclusions:
The performed investigations and theoretical analysis have shown, that the cross-correlation method may be applied in cardiac output measurements. This kind of measurements seems to be more accurate and disturbance-resistant than clinically applied thermodilution.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...