Elastoplasty: first experience in 12 patients.
Thijs A J Urlings1, Edwin van der Linden
1Department of Radiology, Medical Center Westeinde, Postbus 432, 2501, CK, Den Haag, The Netherlands. t_urlings@hotmail.com
Elastoplasty using VK100 silicone cement led to a high rate of silicone pulmonary embolism and perivertebral leakage. This method is not recommended due to significant risks compared to polymethylmethacrylate (PMMA) vertebroplasty.
Area of Science:
- Interventional Radiology
- Biomaterials Science
- Orthopedic Surgery
Background:
- Percutaneous vertebroplasty with polymethylmethacrylate (PMMA) is a common treatment for vertebral compression fractures.
- Restoring vertebral stiffness with PMMA may cause secondary fractures in adjacent vertebrae.
- Elastoplasty is explored as an alternative to potentially prevent these secondary fractures.
Purpose of the Study:
- To retrospectively assess the safety and efficacy of elastoplasty using VK100 silicone cement in patients with vertebral compression fractures.
- To evaluate the incidence of silicone migration, specifically silicone pulmonary embolism and perivertebral leakage.
- To compare the complication rates with those of traditional PMMA vertebroplasty.
Main Methods:
- A retrospective review of 12 patients undergoing elastoplasty between July 2011 and January 2012.
- Exclusion criterion: posterior wall defects.
- Postprocedural chest computed tomography (CT) scans were used to detect perivertebral leakage and pulmonary embolism.
Main Results:
- Twenty-one vertebral bodies were treated in 12 patients.
- Silicone pulmonary emboli were detected in 60% of patients.
- Perivertebral venous plexus leakage occurred in 67% of treated vertebrae.
- One patient experienced severe dyspnea due to multiple peripheral silicone emboli.
Conclusions:
- VK100 silicone cement elastoplasty is associated with a high risk of silicone pulmonary embolism, exceeding that of PMMA.
- The slow setting time of VK100 silicone cement poses a technical challenge for clinical application.
- Current evidence suggests against the use of VK100 silicone cement for elastoplasty.