Low-dose, wide-detector array thoracic aortic CT angiography using an iterative reconstruction technique results in
Prabhakar Rajiah1, Paul Schoenhagen, Dhruv Mehta
1Division of Radiology, Imaging Institute, Cardiovascular Imaging, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH, USA.
Background:
Iterative reconstruction techniques (IRTs) may improve image quality for low-dose imaging compared with filtered back projection (FBP) reconstruction.
Objectives:
We compared the results of an IRT for low-dose thoracic aortic computed tomography (CT) imaging with those from FBP reconstruction.
Methods:
Data from 50 patients who underwent 256-slice CT for evaluation of the thoracic aorta were reconstructed with FBP and an IRT (iDose(4)) at 3 noise-reduction strengths (levels 2, 4, and 6). A blinded reader graded image quality (scale, 1-5; 5 = high diagnostic confidence) and the extent of shoulder artifact (scale, 1-5; 5 = no artifact) on all reconstructions. A second reader evaluated a subset of 20 cases to determine interreader and intrareader reproducibility. The mean and SD of attenuation were measured at 5 locations along the thoracic aorta and both subclavian arteries.
Results:
Image noise (SD of attenuation) improved with IRT relative to FBP (aorta: FBP, 31.4 ± 8.6 HU; IRT level 2, 25.1 ± 6.9 HU; level 4, 21.7 ± 6.2 HU; level 6, 17.2 ± 5.4 HU; P < 0.0001; subclavian arteries: FBP, 92.7 ± 34.6 HU; IRT level 2, 50.1 ± 17.1 HU; level 4, 48.9 ± 18.6 HU; level 6, 45.2 ± 19.2 HU; P < 0.0001), whereas mean attenuation was unchanged. Increasing image quality was observed in the aorta and through the shoulders as the contribution from IRT to the final images increased (P < 0.0001). Significant differences were noted between readers in image quality assessment of the aorta but not through the shoulders.
Conclusion:
IRT is associated with reduced noise and shoulder artifact and allows for low-dose aortic CT imaging.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...