Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor
P Francescon1, W Kilby, N Satariano
1Department of Medical Physics, ULSS6-36100, Vicenza, Italy.
Abstract:
Monte Carlo (MC) simulation of dose to water and dose to detector has been used to calculate the correction factors needed for dose calibration and output factor measurements on the CyberKnife system. Reference field ionization chambers simulated were the PTW 30006, Exradin A12, and NE 2571 Farmer chambers, and small volume chambers PTW 31014 and 31010. Correction factors for Farmer chambers were found to be 0.7%-0.9% larger than those determined from TRS-398 due mainly to the dose gradient across the chamber cavity. For one microchamber where comparison was possible, the factor was 0.5% lower than TRS-398 which is consistent with previous MC simulations of flattening filter free Linacs. Output factor detectors simulated were diode models PTW 60008, 60012, 60017, 60018, Sun Nuclear edge detector, air-filled microchambers Exradin A16 and PTW 31014, and liquid-filled microchamber PTW 31018 microLion. Factors were generated for both fixed and iris collimators. The resulting correction factors differ from unity by up to +11% for air-filled microchambers and -6% for diodes at the smallest field size (5 mm), and tend towards unity with increasing field size (correction factor magnitude <1% for all detectors at field sizes >15 mm). Output factor measurements performed using these detectors with fixed and iris collimators on two different CyberKnife systems showed initial differences between detectors of >15% at 5 mm field size. After correction the measurements on each unit agreed within ∼1.5% at the smallest field size. This paper provides a complete set of correction factors needed to apply a new small field dosimetry formalism to both collimator types on the CyberKnife system using a range of commonly used detectors.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014