Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation
06:54

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation

Published on: August 4, 2023

Electrode migration in cochlear implant patients: not an exception.

Kim S van der Marel1, Berit M Verbist, Jeroen J Briaire

  • 1Department of Otorhinolaryngology, Leiden University Medical Center, Leiden, The Netherlands.

Audiology & Neuro-Otology
|May 26, 2012
PubMed
Summary

Cochlear implant electrode migration occurred in 29% of patients, affecting both those with and without performance complaints. Implant type influenced migration risk, but insertion depth did not.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Radiological risk factors for T2-T3 glottic carcinoma treated with (chemo)radiotherapy-a multicenter analysis.

European radiology·2026
Same author

Preoperative radiological compression features and their relationship with pre- and postoperative visual field defects in pituitary macroadenomas: a retrospective cohort from the neuro-ophthalmological clinic.

Frontiers in ophthalmology·2026
Same author

[<sup>89</sup>Zr]bevacizumab PET/CT imaging of vestibular schwannomas for the prediction of bevacizumab treatment effect in patients with symptomatic <i>NF2</i>-related schwannomatosis: a study protocol for a phase II single centre, prospective, feasibility trial.

BMJ open·2026
Same author

Incorporating softmax in psychophysical detection models for normal and electric hearing.

MethodsX·2026
Same author

The current status and future of using computational models to individually optimise cochlear implant stimulation.

Expert review of medical devices·2026
Same author

Leveraging spatial cues from cochlear implant microphones to efficiently enhance speech separation in naturalistic listening scenes.

Scientific reports·2025

Area of Science:

  • Otolaryngology
  • Biomedical Engineering
  • Neurosurgery

Background:

  • Cochlear implantation is a common treatment for severe to profound hearing loss.
  • Electrode migration can potentially impact device performance and patient outcomes.

Purpose of the Study:

  • To investigate the incidence of cochlear implant electrode migration.
  • To determine if electrode migration correlates with patient complaints or implant type.

Main Methods:

  • Retrospective case review of 35 cochlear implant patients.
  • CT scan analysis for electrode displacement (>1 mm considered migration).
  • Comparison between CII HiFocus1 and HiRes90K HiFocus1J implant types.

Main Results:

Related Experiment Videos

Last Updated: May 22, 2026

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation
06:54

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation

Published on: August 4, 2023

  • Electrode migration detected in 10 patients (29%).
  • HiFocus1 implant type showed significantly less migration compared to HiRes90K HiFocus1J.
  • No correlation found between insertion depth and electrode migration.
  • Two migrations were identified in 5 patients scanned for performance issues.
  • Conclusions:

    • Cochlear implant electrode migration is a notable occurrence in this patient cohort.
    • Migration can happen irrespective of patient-reported symptoms.
    • Implant type is a significant factor in electrode migration risk.