Matrix effects in analysis of β-agonists with LC-MS/MS: influence of analyte concentration, sample source, and SPE
Li-Qi Wang1, Zhen-Ling Zeng, Yi-Juan Su
1National Reference Laboratory of Veterinary Drug Residues (SCAU), College of Veterinary Medicine, South China Agricultural University , Guangzhou 510642, China.
Abstract:
The synergistic influences of analyte concentration, sample source, and solid-phase extraction (SPE) type on matrix effects in the multiresidue analyses of eight β-agonists with LC-ESI-MS/MS were evaluated. Porcine muscle and liver extracts and urine from diverse sources were purified by strong or mixed-mode cation exchange and molecularly imprinted polymer SPE cartridges, respectively. Three spiked concentrations (2, 10, and 20 ng/mL) of eight β-agonists in the purified matrices and the different sample sources were analyzed. The results show that for most β-agonists there are significant differences in matrix effects between analyte concentrations or sample sources (P < 0.05), whereas there is no significant difference in matrix effects between different SPE cartridges (P > 0.05). Results from main effects testing indicated that analyte concentration was the main effector.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometry: Isotope Effect
Mass Analyzers: Overview
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Analyzers: Common Types