Related Experiment Video
Updated: May 21, 2026

Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
Comparison of three buffer solutions for amino acid derivatization and following analysis by liquid chromatography
1Institute of Chemistry, University of Tartu, Ravila 14a, 50411 Tartu, Estonia.
Abstract:
For reversed phase separation amino acids are usually derivatized. Several derivatization reactions are carried out at basic pH. In the present work, influence of three basic buffer solutions on liquid chromatography electrospray ionization mass-spectrometric (LC-ESI-MS) analysis of amino acid derivatives was studied. Borate buffer--the most common derivatization buffer--was found to influence ESI ionization up to 23 min retention time. For 9-fluorenylmethylmethoxycarbonyl chloride (Fmoc-Cl derivatization) carbonate buffer should be preferred as it provides higher responses. Hexafluoroisopropanol (HFIP) buffer improves chromatographic peak shapes and responses for diethyl ethoxymethylenemalonate (DEEMM) derivatives.
Related Concept Videos
Mass Spectrometry: Amine Fragmentation
In amines, the number of nitrogen atoms affects the mass of the molecular ion, which is described by the nitrogen rule of mass spectrometry. This rule states that a compound containing a single or...
Mass Spectrometry of Amines
Peptide Identification Using Tandem Mass Spectrometry