Background:
Current quantitative methods for airborne dichlorodiphenyltrichloroethane (DDT) require collection and extraction times of ≥ 12 h. The aim of this study was to develop a method for quantifying airborne DDT with a short (<4 h) collection and analysis time.
Results:
Precision [relative standard deviation (RSD)] for each calibration point (0.8-9.0), linearity (R(2) = 0.99) and apparent recovery (R' = 96.5%) were determined from thermal desorption (TD) gas chromatography-mass spectrometry (GC-MS) analyses of Tenax-TA-packed sampling tubes spiked with 1-250 ng of DDT. Recovery of (13) C-labeled 4,4'-DDT from tubes spiked before and after air sampling was 97.3 and 90.3% respectively. DDT was detected and quantified in 1-3 L samples of air collected during 10-180 min sampling events. A significant difference was observed in DDT air concentration between 28 and 33 °C during microchamber studies.
Conclusions:
The results demonstrate that the TD GC-MS method developed in this study is precise, reproducible and linear over the span of 1-250 ng of DDT spiked onto TD tubes. By avoiding dilution of the sample, the method described allows the measurement of DDT vapor concentrations during short sampling periods (10-180 min) relevant to mosquito behavior studies.
More Related Videos
07:32Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019
05:20Determination of 45 Pesticides in Avocado Varieties by the QuEChERS Method and Gas Chromatography-Tandem Mass Spectrometry
Published on: December 8, 2023
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Performance Liquid Chromatography: Types of Detectors
Tandem Mass Spectrometry