Controlled laboratory testing of arthroscopic shaver systems: do blades, contact pressure, and speed influence their
Karl Wieser1, Matthias Erschbamer, Stefan Neuhofer
1Orthopaedic Department, University of Zurich, Balgrist Hospital, Zurich, Switzerland. karl.wieser@balgrist.ch
Purpose:
The purposes of this study were (1) to establish a reproducible, standardized testing protocol to evaluate the performance of different shaver systems and blades in a controlled, laboratory setting, and (2) to determine the optimal use of different blades with respect to the influence of contact pressure and speed of blade rotation.
Method:
A holding device was developed for reproducible testing of soft-tissue (tendon and meniscal) resection performance in a submerged environment, after loading of the shaver with interchangeable weights. The Karl Storz Powershaver S2 (Karl Storz, Tuttlingen, Germany), the Stryker Power Shaver System (Stryker, Kalamazoo, MI), and the Dyonics Power Shaver System (Smith & Nephew, Andover, MA) were tested, with different 5.5-mm shaver blades and varied contact pressure and rotation speed. For quality testing, serrated shaver blades were evaluated at 40× image magnification. Overall, more than 150 test cycles were performed.
Results:
No significant differences could be detected between comparable blade types from different manufacturers. Shavers with a serrated inner blade and smooth outer blade performed significantly better than the standard smooth resectors (P < .001). Teeth on the outer layer of the blade did not lead to any further improvement of resection (P = .482). Optimal contact pressure ranged between 6 and 8 N, and optimal speed was found to be 2,000 to 2,500 rpm. Minimal blunting of the shaver blades occurred after soft-tissue resection; however, with bone resection, progressive blunting of the shaver blades was observed.
Conclusions:
Arthroscopic shavers can be tested in a controlled setting. The performance of the tested shaver types appears to be fairly independent of the manufacturer. For tendon resection, a smooth outer blade and serrated inner blade were optimal.
Clinical Relevance:
This is one of the first established independent and quantitative assessments of arthroscopic shaver systems and blades. We believe that this study will assist the surgeon in choosing the optimal tool for the desired effect.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias