Validity of sheet-type portable monitoring device for screening obstructive sleep apnea syndrome
Mina Kobayashi1, Kazuyoshi Namba, Satoru Tsuiki
1Neuropsychiatric Research Institute, Japan Somnology Center, 1-24-10 Yoyogi, Shibuya-ku, Tokyo 151-0053, Japan. kobayashi@somnology.com
Purpose:
The SD-101 is a non-restrictive sheet-like medical device that measures sleep-disordered breathing using pressure sensors that can detect the gravitational alterations in the body that accompany respiratory movement. One report has described that the screening specificity of the SD-101 for mild to moderate obstructive sleep apnea syndrome (OSAS) is relatively low. The present study examines whether the accuracy of the SD-101 for OSAS screening is improved by simultaneously measuring percutaneous oxygen saturation (SpO2).
Methods:
Sixty consecutive individuals with suspected OSAS consented to undergo overnight polysomnography (PSG) together with simultaneous measurements of SD-101 and SpO2 at our laboratory.
Results:
The apnea-hypopnea index (AHI) determined from PSG and the respiratory disturbance index determined from SD-101 measurements significantly correlated (SD-101 alone: r = 0.871, p < 0.0001; SD-101 with SpO2: r = 0.965, p < 0.0001). Bland-Altman plots showed a smaller dispersion for the SD-101 with SpO2 than for the SD-101 alone. The SD-101 with SpO2 detected an AHI of >15 on PSG with a sensitivity and specificity of 96.9 and 90.5 % compared with 87.5 and of 85.7 %, respectively, of the SD-101 alone.
Conclusions:
Simultaneously measuring SpO2 improved the accuracy of the SD-101 for OSAS screening. Furthermore, this modality appears to offer high sensitivity and specificity for detecting even moderately severe OSAS.