Related Experiment Video
Updated: May 21, 2026

Experimental Quantification of Interactions Between Drug Delivery Systems and Cells In Vitro: A Guide for Preclinical Nanomedicine Evaluation
Published on: September 28, 2022
Delivery interaction between co-infused medications: an in vitro modeling study of microinfusion
Amy C Tsao1, Mark A Lovich, Michael J Parker
1Department of Anesthesia, Critical Care and Pain Medicine, Massachusetts General Hospital, Boston, MA 02114, USA.
Objective:
To test the hypothesis that steady-state drug delivery by continuous infusion is predictably affected by a second drug infusion in the same lumen.
Background:
Clinicians commonly administer two drugs by continuous infusion through one central venous catheter lumen (co-infusion). To limit fluid delivery, low flow rate carriers transport concentrated drug solutions; a method called microinfusion. How microinfusion delivery of one drug is affected by a second drug infusion has not been explored.
Methods:
Two water-soluble dyes, tartrazine and erioglaucine, infused at 3 ml · h(-1), modeled drug delivery through a four stopcock linear manifold and catheter lumen. A pump drove a carrier fluid (10 ml · h(-1)). After tartrazine reached steady-state delivery, erioglaucine entered downstream or upstream of the tartrazine infusion. Quantitative spectrophotometry measured dye delivery.
Results:
Starting erioglaucine's infusion upstream of tartrazine's entry caused a transient tartrazine bolus (duration 10 min, peak drug delivery 20% higher than target levels). Starting erioglaucine's infusion downstream produced a similar amplitude, briefer, bolus. Stopping the erioglaucine infusion caused a transient reduction in tartrazine delivery. Measured delivery profiles were comparable to prediction models.
Conclusions:
We confirmed the hypothesis that delivery of one infused drug is transiently affected by starting or stopping a second drug infusion in the same line. The magnitude of the changes can be estimated quantitatively. The clinical impact depends on the drugs being co-infused and patient sensitivity, but could be clinically important; the findings have safety implications for infused medication delivery to critically ill or anesthetized children. We recommend minimizing infusion system dead volumes, connecting the most essential infusion(s) to the main fluid pathway as close as possible to the patient, and recognizing the potential for unintended alterations in delivery when multiple drugs co-infuse.
Related Concept Videos
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Compendial Testing Models II
Methods for Studying Drug Absorption: In vitro
The diffusion cell method uses a two-compartment cell, including a donor compartment with the drug solution, which simulates the environment where the drug is applied, and a receptor compartment with a buffer solution, which simulates the environment...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...