Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy
Published on: August 22, 2018
Inverse Monte Carlo for estimation of scattering and absorption in liquid optical phantoms
Hanna Karlsson1, Ingemar Fredriksson, Marcus Larsson
1Department of Biomedical Engineering, Linköping University, S-581 85 Linköping, Sweden. hanna.karlsson@liu.se
Abstract:
A spectroscopic probe with multiple detecting fibers was used for quantifying absorption and scattering in liquid optical phantoms. The phantoms were mixtures of Intralipid and red and blue food dyes. Intensity calibration for the detecting fibers was undertaken using either a microsphere suspension (absolute calibration) or a uniform detector illumination (relative calibration between detectors). Two different scattering phase functions were used in an inverse Monte Carlo algorithm. Data were evaluated for residual spectra (systematic deviations and magnitude) and accuracy in estimation of scattering and absorption. Spectral fitting was improved by allowing for a 10% intensity relaxation in the optimization algorithm. For a multi-detector setup, non-systematic residual spectrum was only found using the more complex Gegenbauer-kernel phase function. However, the choice of phase function did not influence the accuracy in the estimation of absorption and scattering. Similar estimation accuracy as in the multi-detector setup was also obtained using either two relative calibrated detectors or one absolute calibrated detector at a fiber separation of 0.46 mm.