Coupling of UHPLC with fast fraction collection-microplate scintillation counting and MS for radiolabeled metabolite
Wei Tong1, Swapan Chowdhury, Mark Wrona
1Drug Metabolism & Pharmacokinetics, Merck Research Laboratories, 126 East Lincoln Avenue, Rahway, NJ 07065, USA. wei.tong@merck.com
Background:
To further expand the use of fraction collection (FC)-microplate scintillation counting (MSC) in detecting trace amount of radioactivity in absorption, distribution, metabolism and excretion (ADME) studies and improve the resolution of UHPLC-FC-MSC, we report the coupling of UHPLC with MS and faster FC (1.2 s/fraction) followed by MSC using 384-deep-well LumaPlate™ (PerkinElmer, MA, USA) for profiling of radiolabeled metabolites in plasma, urine, bile and feces.
Results:
Collection of 1.2 s/well clearly improved the resolution of the reconstructed radiochromatograms and, at the same time, provided sufficient detection sensitivity that allowed for more accurate integration of peaks, which is required for radiolabeled ADME studies. The introduction of a reversed gradient as a make-up solvent mixture ensured more uniform drops collected in each well, with resolution maintained throughout the UHPLC run. Less sample injection and more frequent FC resulted in less quenching by matrix and accurate integration of peak.
Conclusion:
UHPLC-FC-MSC-MS is suitable for metabolite profiling in ADME studies and offers higher resolution, higher sensitivity, shorter LC running time, reduced matrix effect and more environmentally friendly experiments compared with conventional online flow scintillation analysis.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: