Related Experiment Video
Updated: May 21, 2026

High Pressure Single Crystal Diffraction at PX^2
Published on: January 16, 2017
Pixellated Cd(Zn)Te high-energy X-ray instrument
1Technology Department, Rutherford Appleton Laboratory, Harwell Science and Innovation Campus, Didcot, OX110QX, U.K.
None:
We have developed a pixellated high energy X-ray detector instrument to be used in a variety of imaging applications. The instrument consists of either a Cadmium Zinc Telluride or Cadmium Telluride (Cd(Zn)Te) detector bump-bonded to a large area ASIC and packaged with a high performance data acquisition system. The 80 by 80 pixels each of 250 microm by 250 microm give better than 1 keV FWHM energy resolution at 59.5 keV and 1.5 keV FWHM at 141 keV, at the same time providing a high speed imaging performance. This system uses a relatively simple wire-bonded interconnection scheme but this is being upgraded to allow multiple modules to be used with very small dead space. The readout system and the novel interconnect technology is described and how the system is performing in several target applications.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
X-ray Imaging
Atomic Emission Spectroscopy: Instrumentation