Revitalizing a vital sign: improving detection of tachypnea at primary triage
William Bianchi1, Andrea Freyer Dugas, Yu-Hsiang Hsieh
1Johns Hopkins University School of Medicine, Department of Emergency Medicine, Baltimore, MD, USA. William.Bianchi@med.lecom.edu
Study Objective:
This study evaluates the accuracy of emergency department (ED) triage respiratory rate measurement using the usual care method and a new electronic respiratory rate sensor (BioHarness, Zephyr Technology Corp.), both compared to a criterion standard measurement.
Methods:
This is a cross-sectional study with convenience sampling conducted in an urban academic adult ED, including 3 separate respiratory rate measurements performed at ED triage: usual care measurement, electronic BioHarness measurement, and criterion standard measurement. The criterion standard measurement used was defined by the World Health Organization as manual observation or auscultation of respirations for 60 seconds. The resultant usual care and BioHarness measurements were compared with the criterion standard, evaluating accuracy (sensitivity and specificity) for detecting tachypnea, as well as potential systematic biases of usual care and BioHarness measurements using a Bland Altman analysis.
Results:
Of 191 analyzed patients, 44 presented with tachypnea (>20 breaths/min). Relative to criterion standard measurement, usual care measurement had a sensitivity of 23% (95% confidence interval [CI] 12% to 37%) and specificity of 99% (95% CI 97% to 100%) for tachypnea, whereas BioHarness had a sensitivity of 91% (95% CI 80% to 97%) and specificity of 97% (95% CI 93% to 99%) for tachypnea. Usual care measurements clustered around respiratory rates of 16 and 18 breaths/min (n=144), with poor agreement with criterion standard measurement. Conversely, BioHarness measurement closely tracked criterion standard values over the range of respiratory rates.
Conclusion:
Current methods of respiratory rate measurement at ED triage are inaccurate. A new electronic respiratory rate sensor, BioHarness, has significantly greater sensitivity for detecting tachypnea versus usual care method of measurement.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Volumes and Capacities I
Acute Respiratory Failure-IV