X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
A Monte Carlo simulation study of a flat-panel X-ray source
Edwin J Grant1, Chrystian M Posada, Carlos H Castaño
1Mining and Nuclear Engineering, Missouri University of Science and Technology, 219 Fulton Hall, 301W 14th Street, Rolla, MO 65409, USA.
Abstract:
A novel flat-panel transmission X-ray source based on nitrogen-incorporated ultra-nano-crystalline diamond (N-UNCD) field emitters is being developed for medical and industrial X-ray imaging. X-ray generation characteristics of the X-ray tubes were simulated with the MCNPX 2.6.0 Monte Carlo code. The simulation results showed that 0.25-12 μm thick tungsten targets could generate an adequate intensity of Bremsstrahlung X-rays. Generated X-ray output intensities of 1.48×10(12)MeV/mA-s of 100 kVp electrons hitting the target were found before collimation.