First application of computed radiology to mammography with synchrotron radiation
1Sincrotrone Trieste S.C.p.A., Basovizza, Trieste, Italy.
Objective:
The aim of this study was to evaluate the feasibility of phase-contrast mammography with synchrotron radiation using a high-resolution computed radiology (CR) system devoted to mammography.
Materials And Methods:
The study was performed at the Synchrotron Radiation for Medical Physics (SYRMEP) beamline of the Elettra synchrotron radiation (SR) facility in Trieste (Italy); X-ray beams were in the range 16-22 keV with a high degree of monochromaticity and spatial coherence. The CR system evaluated is the FCR Profect CS by Fujifilm Global. The first images were obtained from test objects and surgical breast specimens. Images obtained using SR and both screen-film and the CR system were compared with images of the same samples acquired with digital mammography equipment. In view of the good quality of the results obtained, the CR system was used in two mammographic examinations with SR.
Results:
Images acquired using SR and both screen-film and CR were obtained with the same level of delivered dose. Image quality obtained with CR was similar or superior to that of screen-film images. Moreover, the digital images obtained with SR were always better than those acquired using the digital mammography system.
Conclusions:
Phase-contrast mammography with SR using the studied CR system is a feasible option.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation I: X-ray and CT