Precision and accuracy of noninvasive hemoglobin measurements during pregnancy
Eran Hadar1, Oded Raban, Tal Bouganim
1Helen Schneider Hospital for Women, Rabin Medical Center, Petach Tikva and the Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel. dolgaldor@hotmail.com
Introduction:
The NBM-200 is a novel device allowing noninvasive hemoglobin measurement. The system is based on occlusion spectroscopy technology in the red/near-infrared range. At the core of this technology is the production of a new bio-physical signal, resulting from temporarily occluding the blood flow in the measurement site. The measurement is performed using an annular, multi-wavelength probe with pneumatically operated cuffs, with which an over-systolic pressure is produced at the finger base.
Methods:
OrSense NBM200 was tested during the years 2011-2012 in a population of pregnant women. Upon receipt of informed consent, two noninvasive Hemoglobin measurements were performed on the right and left thumbs of each subject. Reference hemoglobin values were obtained from venous blood samples drawn at the same time of the noninvasive measurement. Blood Hemoglobin was evaluated on an LH-750 Beckman Coulter counter, acting as the reference "gold standard."
Results:
A total of 126 data pairs were obtained in the trial from 63 women. The mean error (bias) of the NBM200 readings compared to the reference was 0.1 g/dL and the accuracy, defined as the standard deviation of error, was 0.86 g/dL. A Bland-Altman comparison of the NBM200 versus the Coulter device shows that the 95% limits of agreement is -1.59 to 1.79 g/dL.
Conclusion:
Our study demonstrates a good correlation between reference blood hemoglobin and noninvasive hemoglobin measurements. The NBM-200 can accurately assess hemoglobin levels, in a noninvasive fashion, during pregnancy.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5% chance...