Related Experiment Video
Updated: May 20, 2026

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
A safe overhang limit for unicompartmental knee arthroplasties based on medial collateral ligament strains: an in
Ravindra Gudena1, Mohammad Atarod Pilambaraei, Jason Werle
1Department of Surgery, Faculty of Medicine, University of Calgary, Calgary, Alberta, Canada.
Abstract:
Excessive tibial component overhang during unicompartmental knee arthroplasty (UKA) may cause medial collateral ligament (MCL) impingement, which, in turn, may lead to medial knee pain [Chau et al. Tibial component overhang 226 following unicompartmental knee replacement-does it matter? The Knee. 2009;16(5):310-3]. This study examines MCL loads in 6 human cadaveric knees for different levels of overhang using a robotic testing system. The results indicated no statistically significant difference between the baseline MCL load (no overhang) and the 2-mm overhang (P = .261). However, there were significant differences in MCL load between 2- vs 4-mm (P = .012) and 2- vs 6-mm overhang (P = .022). The loads were almost doubled from 2 to 4 mm of overhang. We conclude that, to minimize pain from excessive MCL loading, surgeons should avoid tibial component overhang greater than 2 mm in unicompartmental knee arthroplasties.