Related Experiment Video
Updated: May 20, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Experimental study of bone response to hydroxyapatite coating implants: bone-implant contact and removal torque test
Tae-Gwan Eom1, Gyeo-Rok Jeon, Chang-Mo Jeong
1Department of Biomedical Engineering, School of Medicine, Pusan National University, South Korea.
Objective:
The objective of this study was to evaluate the early osseointegration of hydroxyapatite (HA)-coated implant.
Study Design:
Twelve adult male miniature pigs were used in this study. The removal torque of implants placed in the tibia of miniature pigs was measured. For implants placed in the mandible, histomorphometric evaluation was performed for the evaluation of the bone-implant contact (BIC) ratio.
Results:
After 4, 8, and 12 weeks, removal torque values were increased. Among the 3 groups, the HA-coated group showed the highest values (P < .05). At 4 and 8 weeks, the BIC ratio of HA was significantly higher than that of resorbable blast medium or sand blasted with alumina and acid etched (P < .05).
Conclusions:
It was concluded that HA-coated implants are relatively favorable in early loading stages.