High-ratio grid considerations in mobile chest radiography.
Alexander W Scott1, David M Gauntt, Michael V Yester
1Department of Radiology, University of Alabama at Birmingham, Birmingham, AL 35249, USA.
High-ratio grids improve mobile chest X-ray image quality without increasing patient dose when aligned using the mobile radiography automatic grid alignment system (MRAGA). Fiber interspace grids generally outperformed aluminum interspace grids.
Area of Science:
- Radiologic Technology
- Medical Imaging Physics
- Diagnostic Radiology
Background:
- Mobile chest radiography often omits grids or uses low-ratio, misaligned grids.
- A novel mobile radiography automatic grid alignment system (MRAGA) ensures accurate focal spot-to-grid alignment.
Purpose of the Study:
- To evaluate the trade-offs of using high-ratio grids in mobile chest radiography.
- To assess image quality at a fixed patient dose with precise focal spot alignment.
Main Methods:
- Utilized modified ANSI chest phantoms (medium and large) with contrast targets.
- Compared five high-ratio grids (10:1-15:1, aluminum and fiber interspace) against nongrid exposures.
- Employed MRAGA for alignment and computed radiography for image acquisition.
- Quantified contrast-to-noise ratio (CNR) and CNR improvement factor (k(CNR)).
Main Results:
- High-ratio grids increased CNR by 4%-65% compared to nongrid images at the same phantom dose.
- Image quality improvement was more pronounced in larger phantoms and mediastinal targets.
- Fiber interspace grids generally yielded superior results over aluminum interspace grids.
- k(CNR) showed minimal dependence on kVp or grid ratio in the lung, but decreased with increasing kVp in the mediastinum.
Conclusions:
- Accurate grid alignment with MRAGA enables high-ratio grids to enhance mobile chest radiography image quality without dose increase.
- Fiber interspace grids demonstrated better performance than aluminum interspace grids.
- Specific fiber grids (13:1 for medium, 15:1 for large phantoms) provided optimal results.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...