Effect of extra-column volume on practical chromatographic parameters of sub-2-μm particle-packed columns in
Naijun Wu1, Ashley C Bradley, Christopher J Welch
1Merck & Co, Inc, Rahway, NJ, USA. nawu@celgene.com
Abstract:
Effects of extra-column volume on apparent separation parameters were studied in ultra-high pressure liquid chromatography with columns and inlet connection tubings of various internal diameters (id) using 50-mm long columns packed with 1.8-μm particles under isocratic conditions. The results showed that apparent retention factors were on average 5, 11, 18, and 41% lower than those corrected with extra-column volumes for 4.6-, 3.0-, 2.1-, and 1.0-mm id columns, respectively, when the extra-column volume (11.3 μL) was kept constant. Also, apparent pressures were 31, 16, 12, and 10% higher than those corrected with pressures from extra-column volumes for 4.6-, 3.0-, 2.1-, and 1.0-mm id columns at the respective optimum flow rate for a typical ultra-high pressure liquid chromatography system. The loss in apparent efficiency increased dramatically from 4.6- to 3.0- to 2.1- to 1.0-mm id columns, less significantly as retention factors increased. The column efficiency was significantly improved as the inlet tubing id was decreased for a given column. The results suggest that maximum ratio of extra-column volume to column void volume should be approximately 1:10 for column porosity more than 0.6 and a retention factor more than 5, where 80% or higher of theoretically predicted efficiency could be achieved.
Related Concept Videos
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...