Investigation into variation and errors of a three-dimensional breast imaging system using multiple stereo cameras
Helga Henseler1, Joanna Smith, Adrian Bowman
1Canniesburn Plastic Surgery Unit Glasgow, Glasgow Royal Infirmary, 84 Castle Street, Glasgow G4 0SF, UK. h.henseler@btinternet.com
Summary
This study found that a 3D multiple stereo camera system for breast assessment showed good reproducibility. While larger breast models had more variable measurements, the system
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Anthropometry
Background:
- Measurement errors in breast volume assessment stem from repeat measurement variability.
- Objective breast assessment requires evaluating the accuracy and precision of measurement systems.
- Understanding error sources is crucial for validating new breast assessment technologies.
Purpose of the Study:
- To evaluate the variation and errors of a three-dimensional (3D) multiple stereo camera system for objective breast assessment.
- To investigate the correlation between measurement errors and breast size.
- To compare the system's performance against established methods like water displacement.
Main Methods:
- Utilized an eight-camera multiple stereophotogrammetry system for 3D imaging.
- Measured volumes of nine plaster breast models and six live volunteers' breasts.
- Assessed variability across different measurement stages and correlated it with model/breast size.
Main Results:
- A significant positive correlation was found between plaster model size and measurement variability (p=0.033).
- A non-significant trend indicated larger live breasts also showed more variability (p=0.342).
- Average variations were 32.95 cc (poses), 19.43 cc (captures), and 28.32 cc (software calculations), representing small proportions of mean volume (4.0%-6.9%).
Conclusions:
- The 3D imaging system demonstrates good reproducibility for breast volume measurement.
- Measurement variation increases with size, a significant finding in models but a trend in live subjects.
- This validation process is essential for adopting new breast measurement technologies.