Related Experiment Videos
Fully automated high-performance liquid chromatographic assay for the analysis of free catecholamines in urine
Abstract:
A totally automated and reliable high-performance liquid chromatographic method is described for the routine determination of free catecholamines (norepinephrine, epinephrine and dopamine) in urine. The catecholamines were isolated from urine samples using small alumina columns. A standard automated method for pH adjustment of urine before the extraction step has been developed. The extraction was performed on an ASPEC (Automatic Sample Preparation with Extraction Columns, Gilson). The eluate was collected in a separate tube and then automatically injected into the chromatographic column. The catecholamines were separated by reversed-phase ion-pair liquid chromatography and quantified by fluorescence detection. No manual intervention was required during the extraction and separation procedure. One sample may be run every 15 min, ca. 96 samples in 24 h. Analytical recoveries for all three catecholamines are 63-87%, and the detection limits are 0.01, 0.01, and 0.03 microM for norepinephrine, epinephrine and dopamine, respectively, which is highly satisfactory for urine. Day-to-day coefficients of variation were less than 10%.