Validating urinary measurement of beta-2-microglobulin with a Roche reagent kit designed for serum measurements
Pak Cheung R Chan1, Vathany Kulasingam, Bonny Lem-Ragosnig
1Department of Clinical Pathology, Sunnybrook Health Sciences Centre, Toronto, ON, M4N 3M5, Canada. PC.Chan@sunnybrook.ca
Objective:
To validate the use of a Roche serum beta-2-microglobulin (B2MG) kit for urinary B2MG measurements, and to establish reference limits for urinary B2MG/creatinine ratio from healthy individuals.
Design And Methods:
The Roche B2MG Tina-Quant serum kit was used to measure urinary B2MG immunoturbidimetrically.
Results:
Using human urine as a diluent, the B2MG method was linear from 73-2156 μg/L. The imprecision on a commercially available urine QC was 4.4% at a concentration of 380 μg/L. Limit of quantification at <20% CV was 40 μg/L. Method comparison with Immulite 2000 (Siemens) yielded slope=1.180 (95% CI 1.14-1.22), intercept=11.5 (95% CI -3.6-26.6), SEE=27.6 and r=0.99 (n=26) by the Deming regression analysis. The upper reference limit of B2MG/creatinine ratio determined from 195 healthy adults was 29 μg/mmol (97.5th centile).
Conclusions:
The serum B2MG Tina Quant reagent kit is acceptable to measure urinary B2MG.