Electromechanical performance of piezoelectric scanning mirrors for medical endoscopy
Kristin H Gilchrist1, David E Dausch, Sonia Grego
1Center for Materials and Electronic Technologies, RTI International, Research Triangle Park, NC, USA.
Abstract:
The electromechanical performance of piezoelectric scanning mirrors for endoscopy imaging is presented. The devices are supported by a single actuating cantilever to achieve a high fill factor, the ratio of mirror area to the combined mirror and actuator area. The largest fill factor devices (74%) achieved 10° mechanical scan range at +/-10V with a 300 microm long cantilever. The largest angular displacement of 30° mechanical scan range was obtained with a 500 microm long cantilever device with a 63% fill factor driven at 40 Vpp. A systematic investigation of device performance (displacement and speed) as a function of fabrication and operational parameters including the stress balance in the cantilever revealed unexpectedly large displacements with lack of inversion at the coercive field. An interpretation of the results is presented based on piezoelectric film domain orientation and clamping with supporting piezoelectric film characterization measurements.