Variation of compound-specific hydrogen isotope ratios under changing temperature program in gas
Weiguo Liu1, Zheng Wang, Linlin Cui
1State Key Laboratory of Loess and Quaternary Geology, IEE, CAS, Xi'an, 710075, PR China. liuwg@loess.llqg.ac.cn
Rationale:
In recent experiments, we found that compound-specific δ(2)H values can vary as a result of changing the gas chromatography temperature program under common pyrolysis conditions. To achieve better precision, it is necessary to examine the details and find a solution to this problem when using gas chromatography/thermal conversion/isotope ratio mass spectrometry (GC-TC-IRMS) for hydrogen isotope analysis.
Methods:
A test was designed to find the possible temperature effect under four different GC temperature ramp rates using n-alkanes (n-C(21), n-C(27), and n-C(31)) and fatty acids (n-C(12), n-C(18), and n-C(24)). The common 'hexane' method was used initially to condition the pyrolysis reactor. Experiments were then carried out using the 'methane condition' method because it was considered to improve pyrolysis efficiency.
Results:
Under the 'hexane condition' the measured hydrogen isotope ratios of the n-alkanes and n-fatty acids became more positive with increasing GC temperature ramp rate. The ion current intensity of hydrogen also generally increased. However, when the 'methane condition' method was used, the measured δ(2)H values of the n-alkanes and n-fatty acids showed little change under different GC temperature ramp rates.
Conclusions:
Higher pyrolysis efficiency could reduce the tailing of the H(2) peak and the related isotopic variations at increased GC temperature ramp rates. In addition, too slow a temperature ramp rate could broaden the peak width and thus increase the background effect and possible isotopic fractionations in the split interface; this could also influence the hydrogen isotope values. We therefore suggest that the appropriate temperature ramp rate is an important factor in improving the precision in analyzing compound-specific hydrogen isotopes.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Isotope Effect
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR
Gas Chromatography: Types of Detectors-II
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...