Reduction of ringing artifacts in the arterial phase of gadoxetic acid-enhanced dynamic MR imaging
Akihiro Tanimoto1, Nobuya Higuchi, Akihisa Ueno
1Department of Diagnostic Radiology, Keio University School of Medicine, Tokyo, Japan. t-mri@tt.rim.or.jp
Purpose:
We assessed what MR imaging parameters affected ringing artifacts during the arterial phase of gadoxetic acid-enhanced dynamic magnetic resonance (MR) imaging.
Materials And Methods:
We tested various parameters, including imaging matrices, choice of either sequential or elliptical centric phase-encoding scheme, scanning time, and contrast injection rate using new simulation software on a personal computer and visually evaluated clinical MR images retrospectively using a 4-point scale to assess ringing artifacts.
Results:
The simulation study revealed that square matrices, short scanning time, slow injection rate, and sequential view ordering effectively reduced ringing artifacts, findings confirmed in clinical practice using gadoxetic acid-enhanced MR imaging. Significantly fewer artifacts resulted using a slow injection rate (P<0.05) and using square matrices in the arterial (P<0.05), portal (P<0.01), and hepatocytic (P<0.05) phases.
Conclusion:
Choice of square matrix, slower injection rate, shorter scanning time, and sequential view ordering could reduce ringing artifacts.
Insights
Optimizing magnetic resonance (MR) imaging for gadoxetic acid-enhanced dynamic scans involves adjusting parameters to reduce ringing artifacts. Using square matrices, shorter scan times, and slower contrast injection rates significantly minimizes these visual disruptions.
Area of Science:
- Radiology
- Medical Imaging
- Magnetic Resonance Imaging
Background:
- Ringing artifacts can degrade image quality in dynamic contrast-enhanced MR imaging.
- Gadoxetic acid-enhanced dynamic MR imaging is crucial for liver lesion characterization.
- Minimizing artifacts is essential for accurate diagnostic interpretation.
Purpose of the Study:
- To identify key magnetic resonance (MR) imaging parameters influencing ringing artifacts.
- To evaluate the impact of different imaging parameters on artifact severity during the arterial phase of gadoxetic acid-enhanced dynamic MR imaging.
Main Methods:
- Utilized novel simulation software for parameter testing.
- Assessed imaging matrices, phase-encoding schemes (sequential vs. elliptical centric), scanning time, and contrast injection rates.
- Retrospectively evaluated clinical MR images using a 4-point scale for artifact assessment.
Main Results:
- Simulation and clinical studies confirmed that square matrices, shorter scanning times, slower contrast injection rates, and sequential view ordering reduce ringing artifacts.
- A slow injection rate significantly decreased artifacts (P<0.05).
- Square matrices demonstrated significant artifact reduction across arterial (P<0.05), portal (P<0.01), and hepatocytic (P<0.05) phases.
Conclusions:
- Square matrix selection is effective in reducing ringing artifacts.
- Slower contrast injection rates and shorter scanning times contribute to artifact mitigation.
- Sequential view ordering can also help minimize ringing artifacts in gadoxetic acid-enhanced dynamic MR imaging.
Related Concept Videos
Magnetic Resonance Imaging
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...