Related Experiment Video
Updated: May 20, 2026

Rapid Quantification of Oxidized and Reduced Forms of Glutathione Using Ortho -phthalaldehyde in Cultured Mammalian Cells In Vitro
Published on: June 28, 2024
A simple microfluidic integrated with an optical sensor for micro flow injection colorimetric determination of
Sam-ang Supharoek1, Napaporn Youngvises, Jaroon Jakmunee
1Department of Chemistry, and Center of Excellence for Innovation in Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai, Thailand.
Abstract:
A simple and inexpensive method for fabricating a microfluidic platform was developed. A printed circuit board (PCB) was used to make a master mold for replicating a polydimethylsiloxane (PDMS) microchannel. The master mold was fabricated by a simple photolithographic method, employing a photoresist dry film. The process did not use hazardous chemicals, a clean room or any expensive instrument. The PDMS microchannel was clamped with polymethylmethacrylate (PMMA) plates, where a light emitting diode (LED) as a light source and a light dependent resistor (LDR) as a light sensor were attached to form a simple optical sensor. The system was successfully employed as a micro flow injection analysis for the determination of glutathione in dietary supplement samples. A linear calibration graph in the range of 5.0 - 60.0 mg L(-1) glutathione was obtained with a detection limit of 0.01 mg L(-1). The system provided a sample throughput of 48 h(-1), with microliter consumption of the reagent.