Related Experiment Video
Updated: May 20, 2026

Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Validation study and routine control monitoring of moist heat sterilization procedures
1School of Science, Chuo University, Tokyo, Japan. shintani@mail.hinocatv.ne.jp
Abstract:
The proposed approach to validation of steam sterilization in autoclaves follows the basic life cycle concepts applicable to all validation programs. Understand the function of sterilization process, develop and understand the cycles to carry out the process, and define a suitable test or series of tests to confirm that the function of the process is suitably ensured by the structure provided. Sterilization of product and components and parts that come in direct contact with sterilized product is the most critical of pharmaceutical processes. Consequently, this process requires a most rigorous and detailed approach to validation. An understanding of the process requires a basic understanding of microbial death, the parameters that facilitate that death, the accepted definition of sterility, and the relationship between the definition and sterilization parameters. Autoclaves and support systems need to be designed, installed, and qualified in a manner that ensures their continued reliability. Lastly, the test program must be complete and definitive. In this paper, in addition to validation study, documentation of IQ, OQ and PQ concretely were described.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Temperature
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Bioreactor Controls-I
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration