Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Multiple Comparison Tests01:13

Multiple Comparison Tests

Multiple comparison test, abbreviated as MCT, is a post hoc analysis generally performed after comparing multiple samples with one or more tests. An MCT will help identify a significantly different sample among multiple samples or a factor among multiple factors.
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Quantifying Heat02:46

Quantifying Heat

Thermal Energy Microscopically, thermal energy is the kinetic energy associated with the random motion of atoms and molecules. Temperature is a quantitative measure of “hot” or “cold”, which depends on the amount of thermal energy. When the atoms and molecules in an object are moving or vibrating quickly, they have a higher average kinetic energy (KE) (or higher thermal energy), and the object is perceived as “hot”, or it is described as being at a higher temperature. When the atoms and...
Physical Methods for Controlling Microbial Growth: Temperature01:23

Physical Methods for Controlling Microbial Growth: Temperature

Heat is a widely used method to control microbial growth by targeting and denaturing cellular proteins, thereby killing or inactivating microbes. This method's effectiveness is quantified using parameters such as the thermal death point (TDP), thermal death time (TDT), and decimal reduction time (D value). TDP represents the lowest temperature at which all microorganisms in a liquid suspension are eliminated within 10 minutes, whereas TDT is the time necessary to achieve sterilization at a...
Calorimetry01:19

Calorimetry

When objects at different temperatures are placed in contact with each other but isolated from everything else, they attain thermal equilibrium. A container that prevents heat transfer in or out is called a calorimeter, and the use of a calorimeter to make measurements is called calorimetry. Generally, these measurements involve heat or specific heat capacity. The term "calorimetry problem" is used for any problem where the specified objects are thermally isolated from their surroundings. An...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

In Vitro Comparison of Porcelain Fused to Metal Crown Retention after Endodontic Access and Subsequent Restoration: Composite, Amalgam, Amalgam with Composite Veneer, and Fiber Post with Composite.

Journal of endodontics·2020
Same author

Relationship between Canal Enlargement and Fracture Load of Root Dentin Sections.

Dental materials : official publication of the Academy of Dental Materials·2019
Same author

Mandibular Canal Location: Cone-beam Computed Tomography Examination.

Journal of endodontics·2016
Same author

Cone-beam Computed Tomography: Anatomic Analysis of Maxillary Posterior Teeth-Impact on Endodontic Microsurgery.

Journal of endodontics·2016
Same author

Dens Evaginatus: Just Another Cusp?

Northwest dentistry·2015
Same author

Electrical Pulp Testing: Sources of Error.

Northwest dentistry·2015
Same journal

Clinical Outcome of Replanted Avulsed Anterior Teeth: A Retrospective Cohort Study.

Journal of endodontics·2026
Same journal

Mahidol study 3: Post-operative Effect and Adverse Events of Regenerative Endodontic Procedures in Immature Permanent Teeth.

Journal of endodontics·2026
Same journal

Non-Surgical Management of Horizontal Root Fracture with Necrotic Apical Segment and Coronal Cervical Abfraction Using Mineral Trioxide Aggregate: Case Report.

Journal of endodontics·2026
Same journal

Identification of pain-related biomarkers and the associated potential molecular regulation mechanism in pulpitis by bioinformatics method.

Journal of endodontics·2026
Same journal

AI Prognostic Models in Endodontics: Scoping Review and Clinical Gap Analysis for Retreatment vs Microsurgery Decision Support.

Journal of endodontics·2026
Same journal

AI-based 3D measurement of root canal curvature from CBCT: Validation of an automated Schneider angle analysis.

Journal of endodontics·2026
See all related articles

Related Experiment Video

Updated: May 20, 2026

High-Throughput Assays of Critical Thermal Limits in Insects
06:58

High-Throughput Assays of Critical Thermal Limits in Insects

Published on: June 15, 2020

Comparison of heat-testing methodology.

Mark M Bierma1, Scott McClanahan, Michael K Baisden

  • 1Division of Endodontics, University of Minnesota School of Dentistry, Minneapolis, MN 55455, USA.

Journal of Endodontics
|July 17, 2012
PubMed
Summary
This summary is machine-generated.

The Elements unit with a heat-testing tip offers the most reliable pulp warming for diagnosing irreversible pulpitis. This method ensures consistent temperature increases, aiding clinicians in differentiating between normal and inflamed dental pulps safely.

More Related Videos

Heat Tolerance Assays Using the Drosophila Activity Monitor System: A Guide to an Executable Application for Data Analysis
05:05

Heat Tolerance Assays Using the Drosophila Activity Monitor System: A Guide to an Executable Application for Data Analysis

Published on: December 13, 2024

Related Experiment Videos

Last Updated: May 20, 2026

High-Throughput Assays of Critical Thermal Limits in Insects
06:58

High-Throughput Assays of Critical Thermal Limits in Insects

Published on: June 15, 2020

Heat Tolerance Assays Using the Drosophila Activity Monitor System: A Guide to an Executable Application for Data Analysis
05:05

Heat Tolerance Assays Using the Drosophila Activity Monitor System: A Guide to an Executable Application for Data Analysis

Published on: December 13, 2024

Area of Science:

  • Endodontics
  • Dental Diagnostics

Background:

  • Diagnosing irreversible pulpitis often involves heat sensitivity tests.
  • Current heat tests can cause damaging pulp temperature increases.
  • Newer electronic devices offer controlled pulp warming.

Purpose of the Study:

  • To evaluate the consistency and safety of different pulp warming methods.
  • To identify the optimal technique for reproducing heat sensitivity in irreversible pulpitis diagnosis.

Main Methods:

  • Four heat application methods were tested on extracted premolars: heated gutta-percha, heated ball burnisher, hot water, and an electronic heat-testing tip.
  • Temperatures were recorded in the pulp chamber over 60 seconds.
  • Intraclass correlation coefficient was used for data analysis.

Main Results:

  • Hot water showed the least consistent warming.
  • The electronic heat-testing tip provided superior consistency among operators.
  • Hot water and heated ball burnisher generated potentially damaging temperatures.

Conclusions:

  • The electronic heat-testing tip ensures the most consistent dental pulp warming.
  • This method aids in reliable differentiation between normal and irreversible pulpitis.