The effects of target motion in kV-CBCT imaging
Sriram Padmanaban1, Raghavendiran Boopathy, Bhuvana Kunjithapatham
1Cancer Institution (WIA), Adyar Chennai Tamilnadu, India.
Background:
To study the impact of target motion in kV-CBCT imaging.
Material/Methods:
To simulate the respiratory movement, dynamic phantom was programmed to move in three-dimension with a period of four seconds and of two different amplitudes (PA1 and PA2). The targets of well defined geometries (cylinder, sphere, solid triangle, U-shaped and dumbbell) were made using wax. The static targets were CT imaged (reference image). Using CBCT, the targets in static and dynamic modes were imaged under full-fan beam. The line profiles along cranio-caudal direction, influence of target's initial moving phase and volume estimation using auto-contouring tool were used to analyze the effects of target motion on CBCT imaging.
Results:
Comparing the line profiles of targets in CBCT with CT, the length of average HU spread was reduced by 42.54±1.85%, except the cylindrical target which is by 19.35% for PA1. The percentage difference in reconstructed volume of static targets imaged using CBCT and CT (HU WW -500 to 0) ranges from -1.32% to -5.94%. The volume losses for targets imaged in dynamic mode PA1 ranges from 14.35% to 30.95% and for PA2 it was 21.29% to 43.80%. The solid triangle and cylindrical targets suffered the maximum and minimum volume losses respectively.
Conclusions:
Non-gated CBCT imaging of the moving targets encounters significant loss of volumetric information, due to scatter artifacts. These may result in a systematic error in re-contouring when CBCT images are used for the re-planning work.
More Related Videos
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
07:03Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration
Published on: February 23, 2017
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...