[Evaluation of commercial resolution recovery techniques in four state-of-the-art single photon emission computed
Hideo Onishi1, Nobutoku Motomura, Koichi Fujino
1Graduate School of Comprehensive Scientific Research, Prefectural University of Hiroshima, Japan.
Purpose:
This study aims to evaluate the fundamental performance of four leading advanced resolution recovery methods.
Method:
To evaluate the performance of the resolution recovery algorithm, we carried out the computer simulation with the cone/sphere digital phantoms. These phantoms were used to investigate the basic properties of those algorithms. The software of four packages (advance) were also tested, specifically Astonish(TM) (AST), Evolution(TM) (EVL), Flash-3D(TM) (FL3), and 3D-OSEM (3DOS). The performance was evaluated in the collimator systems (LEHR) reconstruction conditions using the full width at half maxi am (FWHM), aspect ratio (ASR), and artifacts of conical part.
Result:
In the "without BG," FWHM of the advance method indicated a true-FWHM with SI (subset×iteration)=20, 40. As SI increased, FWHM was composed with over estimate. Each advances of FWHM indicated only 5% of improvement as compared with reference FWHM in the "with BG." The ASR increased 20% to AST, FL3, and ASR of 3DOS remained in 10% in the outside. As for the reproducibility of the conical part, an artifact was caused by the FL3, EVL, and AST methods. This artifact did not occur in 3DOS.
Discussions:
An SI needs more than 150 to obtain an accurate compensation effect. As for the advance method, the major compensation effect was not demonstrated very much as compared with the OS-EM. The EVL, FL3, and AST overestimated values due to a Gibb's oscillation in the artifacts of the conical part.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...