Improved quantification of radionuclide uptake using deconvolution and windowed subtraction techniques for scatter
J C Yanch1, M A Flower, S Webb
1Joint Department of Physics, Institute of Cancer Research, Sutton, Surrey, England.
Abstract:
A comparison of two methods of scatter compensation in single photon emission computed tomography (SPECT) imaging is made on the basis of improvements in quantification. The methods, scatter-window subtraction and constrained deconvolution of an average point source response function (PSRF), are described; the theoretical basis of each method is also briefly assessed. Improvements in relative quantification offered by each method are measured by examining ratios of counts in hot cylinders to counts in a slightly radioactive background. The cylinder/background concentration ratio was varied by a factor of five; the sizes of the cylinders remained constant. Keeping the diameter of the cylinders constant allowed for an assessment of the effect of concentration differences, without the conflicting effect of variation in the size of the hot source. Results showed that while both scatter-window subtraction and constrained deconvolution offer quantification improvements in the final image, the method of prereconstruction deconvolution of a planar PSRF from each planar projection is substantially more successful than either scatter-window subtraction or other methods of implementing the deconvolution procedure.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET