Comparison of dose decrement from intrafraction motion for prone and supine prostate radiotherapy
Jeffrey R Olsen1, Parag J Parikh, Michael Watts
1Department of Radiation Oncology, Washington University School of Medicine, St. Louis, MO 63110, USA.
Background And Purpose:
Dose effects of intrafraction motion during prone prostate radiotherapy are unknown. We compared prone and supine treatment using real-time tracking data to model dose coverage.
Material And Methods:
Electromagnetic tracking data were analyzed for 10 patients treated prone, and 15 treated supine, with IMRT for localized prostate cancer. Plans were generated using 0 mm, 3 mm, and 5mm PTV expansions. Manual beam-hold interventions were applied to reposition the patient when translations exceeded a predetermined threshold. A custom software application (SWIFTER) used intrafraction tracking data acquired during beam-on model delivered prostate dose, by applying rigid body transformations to the prostate structure contoured at simulation within the planned dose cloud. The delivered minimum prostate dose as a percentage of planned dose (Dmin%), and prostate volume covered by the prescription dose as a percentage of the planned volume (VRx%) were compared for prone and supine treatment.
Results:
Dmin% was reduced for prone treatment for 0 (p=0.02) and 3 mm (p=0.03) PTV margins. VRx% was reduced for prone treatment only for 0mm margins (p=0.002). No significant differences were found using 5 mm margins.
Conclusions:
Intrafraction motion has a greater impact on target coverage for prone compared to supine prostate radiotherapy. PTV margins of 3 mm or less correlate with a significant decrease in delivered dose for prone treatment.