Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
Isocentric 3-dimensional C-arm imaging of component alignments in total knee arthroplasty with potential
Shahram Amiri1, David R Wilson, Carolyn Anglin
1Department of Orthopaedics, University of British Columbia, Vancouver, British Columbia, Canada.
Abstract:
An intraoperative imaging tool for total knee arthroplasty could help avoid poor clinical outcomes related to malalignment. We investigated the feasibility of using isocentric (ISO-C) fluoroscopic imaging for this purpose. Three-dimensional ISO-C and computed tomographic (CT) images were acquired from 6 cadaveric specimens implanted with standard knee arthroplasty components and analyzed to determine rotational alignments. In comparison with standard CT measures, the ISO-C-based measures had overall accuracies (determined as root mean square error) of 0.8° and 1.3° and corresponding SDs of 1.3° and 1.4° for the femoral and tibial components, respectively. With ISO-C imaging, it is possible to measure rotational alignment in knee arthroplasty with accuracy and repeatability comparable with CT. Isocentric imaging has strong potential as an intraoperative tool to accurately align arthroplasty components.