Acute hyperkalemia detected by alert from implantable cardioverter-defibrillator
Vikas Kuriachan1, Usha Tedrow, Elliott Antman
1Cardiovascular Division, Department of Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA. vikas.kuriachan@albertahealthservices.ca
Abstract:
Two cases of implantable cardioverter-defibrillator lead impedance changes that corresponded with changes in serum potassium are described. In the first case, an audible lead alert was triggered due to a significant change in lead impedance caused by hyperkalemia. In the second case, a patient was admitted with hyperkalemia and renal failure. Device interrogation revealed a change in lead impedance corresponding to changes in serum potassium. As devices and leads become more sophisticated, their role for monitoring parameters and triggering alerts may expand to include electrolyte and other metabolic changes.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias V: Evaluating Dysrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Cardiomyopathy II: Dilated Cardiomyopathy