Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Advanced material modelling in numerical simulation of primary acetabular press-fit cup stability
R Souffrant1, C Zietz, A Fritsche
1Biomechanics and Implant Technology Research Laboratory, Department of Orthopaedics, University of Rostock, Rostock, Germany. robert.souffrant@med.uni-rostock.de
Abstract:
Primary stability of artificial acetabular cups, used for total hip arthroplasty, is required for the subsequent osteointegration and good long-term clinical results of the implant. Although closed-cell polymer foams represent an adequate bone substitute in experimental studies investigating primary stability, correct numerical modelling of this material depends on the parameter selection. Material parameters necessary for crushable foam plasticity behaviour were originated from numerical simulations matched with experimental tests of the polymethacrylimide raw material. Experimental primary stability tests of acetabular press-fit cups consisting of static shell assembly with consecutively pull-out and lever-out testing were subsequently simulated using finite element analysis. Identified and optimised parameters allowed the accurate numerical reproduction of the raw material tests. Correlation between experimental tests and the numerical simulation of primary implant stability depended on the value of interference fit. However, the validated material model provides the opportunity for subsequent parametric numerical studies.