Related Experiment Video
Updated: May 20, 2026

Ecotoxicological Methodologies to Evaluate Biomarkers at Different Scales in Neotropical Anurans
Published on: April 28, 2023
Flow injection catalase activity measurement based on gold nanoparticles/carbon nanotubes modified glassy carbon
1Faculty of Pharmacy and Biotechnology, The German University in Cairo, New Cairo City, Egypt. rasha.elnashar@guc.edu.eg
Abstract:
Amperometric flow injection method of hydrogen peroxide analysis was developed based on catalase enzyme (CAT) immobilization on a glassy carbon electrode (GC) modified with electrochemically deposited gold nanoparticles on a multiwalled carbon nanotubes/chitosan film. The resulting biosensor was applied to detect hydrogen peroxide with a linear response range 1.0×10(-7)-2.5×10(-3)M with a correlation coefficient 0.998 and response time less than 10s. The optimum conditions of film deposition such as potential applied, deposition time and pH were tested and the flow injection conditions were optimized to be: flow rate of 3ml/min, sample volume 75μl and saline phosphate buffer of pH 6.89. Catalase enzyme activity was successfully determined in liver homogenate samples of rats, raised under controlled dietary plan, using a flow injection analysis system involving the developed biosensor simultaneously with spectrophotometric detection, which is the common method of enzymatic assay.