Theoretical comparison of the performance of gradient elution chromatography at constant pressure and constant flow
Fabrice Gritti1, Georges Guiochon
1Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:
The theory of gradient elution usually assumes a constant flow rate. This works extends it to gradients performed under constant pressure drop and variable flow rates. The peak capacity is derived under both constant flow rate and constant pressure gradient chromatography by integrating the rate of increase of the peak capacity from the hold-up time to the end of the gradient time. Assuming that the eluent mixture is incompressible, the chromatographic system isothermal, the pressure has no effect on the retention pattern, and neglecting the contributions of the instrument to the total pressure drop and the total peak width, it is found that both modes of gradient elution chromatography are strictly equivalent, provided that the elution time of the last eluted compound and the volume gradient are kept the same in both cases.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Chromatographic Methods: Terminology