Related Experiment Video
Updated: Aug 9, 2026

Biochemical Reconstitution of Steroid Receptor•Hsp90 Protein Complexes and Reactivation of Ligand Binding
Published on: September 21, 2011
A robust assay for dexamethasone in plasma using a heterologous 125I radioligand and a magnetizable solid-phase
N J Cook1, G F Read, C Lejeune-Lenain
1Tenovus Institute for Cancer Research, University of Wales College of Medicine, Heath Park, Cardiff, UK.
Abstract:
An assay using an antiserum raised against a dexamethasone 21-hemisuccinate conjugate and the heterologous radioligand dexamethasone 21-(carboxymethyl) ether was developed, validated, and used to study the pharmacokinetics of this steroid for 12 h following administration to patients with congenital adrenal hyperplasia. Coupling the antiserum to magnetizable cellulose allowed rapid separation of bound/free steroid. A C-21 rather than a C-3 antiserum was used to minimize interference with a main metabolite, 6 beta-hydroxydexamethasone. Close correspondence of assay (0.35 nmol/L) and curve (0.25 nmol/L) sensitivities suggests that interference by matrix effects is minimal. This was confirmed by good agreement in data from the in-house assay and that of a reference procedure. Good precision was demonstrated by the precision profile and Shewhart chart quality control data. The latter also demonstrated the assay was robust and reliable in routine practice.